ChipFind - документация

Электронный компонент: LPS65-M

Скачать:  PDF   ZIP
Technical Support: (888) 41-ASTEC or (407) 241-2752
Americas: (760) 930-4600 Europe (UK) 44 (1384) 842-211
Asia (HK) 852-2437-9662
94
Special Features
60 Watts
LPS60 Series
Medical
Electrical Specs
rev 08.11.04
VDE
0750/EN60601-1 (IEC601)
21310-3336-0015
UL
UL2601
E182560
CSA
CSA 22.2 No. 601.1
LR109492
CE
Mark (LVD)
Medical safety approvals
Universal input
3" x 5" footprint
Remote sense
Built-in EMI filter
Low output ripple
Adjustable output
Overvoltage protection
Overload protection
110 KHz switching frequency
LPX60 enclosure kit available
Operating temperature: 0 to 50C ambient
derate each output at 2.5% per degree from 50 to
70C
Electromagnetic susceptibility: designed to meet
IEC 801, -2, -3, -4, -5, -6, Level 3
Humidity: Operating; non-condensing
5% to 95%
Vibration: Three orthogonal axes, sweep at 1
oct/min, 5 min. dwell at four major resonances
0.75 G peak 5 Hz to 500 Hz, operational
Storage temperature: -40 to 85C
Temperature coefficient: .04% per C
MTBF demonstrated: >550,000 hours at full load
and 25C ambient conditions
Input
Input range
85-264 VAC
120-300 VDC
Frequency
47-440 Hz
Inrush current
<18 A peak @ 115 VAC;
<36 A peak @ 230 VAC,
cold start @ 25C
Input current
1.5 A max. (RMS) @ 115 VAC
Efficiency
70% typical at full load
EMI filter
FCC Class A conducted
CISPR 22 Class A conducted
EN55022 class A conducted
VDE 0878 PT3 class A conducted
Safety ground leakage
current
<75 A @ 50/60 Hz, 264 VAC input
Output
Maximum power
60 W for convection;
80 W with 30 CFM forced air
Adjustment range
-5, +10% minimum
Hold-up time
20 ms @ 60 W load, 115 VAC nominal line
Overload protection
Short circuit protection on all outputs.
Case overload
protected @ 110-145%
above peak rating
Overvoltage protection
5 V output: 5.7-6.7 VDC. Other outputs
10% to 25% above nominal output
Remote sense
Compensates for 0.5 V lead drop min. Will
operate without remote sense connected.
Reverse connection protected.
Environmental
Safety
Total Power:
60-80 Watts
Input Voltage: 85-264 VAC
120-300 VDC
# of Outputs:
Single
EUROPE
Astec House, Waterfront Business Park
Merry Hill, Dudley
West Midlands, DY5 1LX, UK
Telephone: 44 (1384) 842-211
Facsimile: 44(1384) 843-355
AMERICAS
5810 Van Allen Way
Carlsbad, CA 92008
Telephone: 760-930-4600
Facsimile: 760-930-0698
95
www.astecpower.com
Ordering Information
ASIA
Units 2111-2116, Level 21
Tower1, Metroplaza
223, Hing Fong Road
Fwai Fong, New Territories
Hong Kong
Telephone: 852-2437-9662
Facsimile: 852-2402-4426
Model Output
Minimum
Maximum Load with
Maximum Load with
Peak
Regulation2
Ripple
Number
Voltage
Load
Convection Cooling
30 CFM Forced Air
Load1
P/P (PARD)3
LPS63-M
12 V
0 A
5 A
6.7 A
7.5 A
2%
120 mV
LPS64-M
15 V
0 A
4 A
5.3 A
6 A
2%
150 mV
LPS65-M
24 V
0 A
2.5 A
3.3 A
3.8 A
2%
240 mV
1. Peak current lasting <30 seconds with a maximum 10% duty cycle.
2. At 25C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings.
3. Peak-to-peak with 20 MHz bandwidth and 10 F in parallel with a 0.1 F capacitor at rated line voltage and load ranges.
Connector
LPS63-M
LPS64-M
LPS65-M
SK1-1
Neutral
Neutral
Neutral
SK1-3
Line
Line
Line
SK2-1
+12 V
+15 V
+24 V
SK2-2
+12 V
+15 V
+24 V
SK2-3
+12 V
+15 V
+24 V
SK2-4
Common
Common
Common
SK2-5
Common
Common
Common
SK2-6
Common
Common
Common
SK201-1
+Sense
+Sense
+Sense
SK201-2
-Sense
-Sense
-Sense
Mating Connectors
AC Input:
Molex 09-50-8031 (USA)
09-91-0300 (UK)
PINS: 08-58-0111
DC Outputs:
Molex 09-50-8061 (USA)
09-91-0600 (UK)
PINS: 08-58-0111
Remote Sense: Molex 22-01-2025
PINS: 08-50-0114
Astec Connector Kit #70-841-006, includes all of the above.
Notes:
1. Specifications subject to change without notice.
2. All dimensions in inches (mm), tolerance is .02".
3. Mounting holes M1 and M2 should be grounded for EMI purposes.
4. Mounting hole M1 is safety ground connection.
5. Specifications are for convection rating at factory settings at 115 VAC
input, 25C unless otherwise stated.
6. Warranty: 1 year
7. Weight: 0.75 lb. / 0.34 kg
Pin Assignments
3.00"
(76.2)
2.55"
(64.8)
0.22"
(5.6)
0.22"
(5.6)
4.56"
(115.8)
5.00"
(127)
1.65"
(41.9)
0.12"
(3)
SK 1
SK 2
1
2
3
4
5
6
0.15" (4)
DIA (4 places)
GND
1
3
N
L
M1
M2
Remote Sense
Connector
_
+
SK201
Voltage
Adj. Pot
Power Derating Curve
Output
Power
(Watts)
0C
20W
40W
60W
80W
10C 20C 30C 40C 50C 60C 70C
TA Ambient Temperature C
0W
Natural
Convection
Cooling
40W
30W
30 CFM Forced Air Cooling
LPS60 Medical Series